QPoint® Tap Water Filter Assembly product photo Primary L

QPoint® Tap Water Filter Assembly

Combining Fixed Docking Station and Exchangeable Filter Capsule for up to 2 Months Use

The QPoint Tap Water Filter Assembly with exchangeable Filter Capsule for up to 62 day (2 months) use provides 0.2 μm sterilizing-grade filtered water at the Point-of-Use.

The Filter Capsules have been validated according to ASTM F838-20 and demonstrated complete retention as per FDA requirements. In addition, these Filter Capsules have undergone multiple intermittent hot water cycles at 5 bar (75 psi), followed by testing according to ASTM F838-20 with complete retention of Brevundimonas diminuta as per FDA, thus meeting operating conditions as defined in HIMA. Furthermore, field evaluations confirm performance under real life use conditions.

The high technology integrated prefiltration layer delivers exceptional particulate trapping capacity with good flow rates, and protects the proven double layer sterilizing grade membranes enabling long performance life. The Filter Capsule retains fungi, protozoa and bacteria such as Legionella spp., Pseudomonas spp., non-tuberculous Mycobacteria and Escherichia coli without volume restriction. A combination of the technologies within the Filter Capsule minimizes the risk of retrograde contamination over the life of the product.

The smooth, chrome finish Docking Station is hygienically designed, and will fit to all standard tap water outlets. The combination of this fixed Docking Station with the disposable Filter Capsule is compact, reduces waste, reduces costs and improves logistics for the user.
 
Features Benefits
Double layer 0.2 µm sterilizing grade filter
membranes
Retains microorganisms without volume restriction
Fixed Docking Station and exchangeable Filter Capsule Waste reduction, cost reduction and improved logistics
Robust pre-filtration Higher dirt capacity for better flow rates
Compatible with commonly used systemic disinfection Integrates with water safety plans
 


Need Help?
Ask a Question
* Required

Description

Combining Fixed Docking Station and Exchangeable Filter Capsule for up to 2 Months Use

The QPoint Tap Water Filter Assembly with exchangeable Filter Capsule for up to 62 day (2 months) use provides 0.2 μm sterilizing-grade filtered water at the Point-of-Use.

The Filter Capsules have been validated according to ASTM F838-20 and demonstrated complete retention as per FDA requirements. In addition, these Filter Capsules have undergone multiple intermittent hot water cycles at 5 bar (75 psi), followed by testing according to ASTM F838-20 with complete retention of Brevundimonas diminuta as per FDA, thus meeting operating conditions as defined in HIMA. Furthermore, field evaluations confirm performance under real life use conditions.

The high technology integrated prefiltration layer delivers exceptional particulate trapping capacity with good flow rates, and protects the proven double layer sterilizing grade membranes enabling long performance life. The Filter Capsule retains fungi, protozoa and bacteria such as Legionella spp., Pseudomonas spp., non-tuberculous Mycobacteria and Escherichia coli without volume restriction. A combination of the technologies within the Filter Capsule minimizes the risk of retrograde contamination over the life of the product.

The smooth, chrome finish Docking Station is hygienically designed, and will fit to all standard tap water outlets. The combination of this fixed Docking Station with the disposable Filter Capsule is compact, reduces waste, reduces costs and improves logistics for the user.
 
Features Benefits
Double layer 0.2 µm sterilizing grade filter
membranes
Retains microorganisms without volume restriction
Fixed Docking Station and exchangeable Filter Capsule Waste reduction, cost reduction and improved logistics
Robust pre-filtration Higher dirt capacity for better flow rates
Compatible with commonly used systemic disinfection Integrates with water safety plans
 

Specifications

QPoint Docking Station - Tap Assembly (QDTC)

 
Diameter (nominal) 88 mm (3.5 in.)
Weight (nominal) 84 g (3.0 oz)
Temperature Continuous influent temperature 60 °C (140 °F)
Maximum influent temperature 75 °C (167 °F) for a total cumulative period of 90 minutes over the life of each Filter Capsule
Maximum Operating Pressure 5 bar (75 psi) at continuous 60 °C (140 °F) influent temperature
Cleansing and Disinfection External surfaces may be cleaned using common sanitising agents
During routine water system disinfection the Docking Station with attached Filter Capsule can be left in place to enable the outlet to be flushed
Main Materials of Construction Polycarbonate/Acrylonitrile Butadiene Styrene (51%), Glass Filled Nylon (22%), Polyester (20%), Stainless Steel (4%)
Packaging Recycled cardboard outer packaging
 

QPoint Filter Capsule (QJ212, QJ22)

 
Flow Rates (nominal) 5.7 L / min at 1 bar
13.1 L / min at 3 bar
18.8 L / min at 5 bar
(1.5 gal / min at 15 psi)
(3.5 gal / min at 44 psi)
(5.0 gal / min at 75 psi)
Diameter (nominal) 88 mm (3.5 in.)
Weight (nominal) 119 g (4.2 oz)
Temperature Continuous influent temperature 60 °C (140 °F)
Maximum influent temperature 75 °C (167 °F) for a total cumulative period of 90 minutes over the life of each Filter Capsule
Maximum Operating Pressure 5 bar (75 psi) at continuous 60 °C (140 °F) influent temperature
Sanitisation Compatible and complementary to common surface and systemic disinfection programs
Main Materials of Construction Polyester (83.4%), Polypropylene (10.1%), Polyethersulphone (6.5%), (Recyclable)
Packaging Recycled cardboard outer packaging. Ethylene Vinyl Acete (EVA) plastic film blister. Tyvek polyethylene lid
 

Type

Point-of-Use Filters

Additional Information

Pall QPoint& Docking Station (QDTC)

Pictorial Installation Guide

Description

The Pall QPoint Docking Station - Tap Assembly with Quick Connect Inlet is supplied non-sterile.

For full details, refer to the Instructions for Use.
  • Remove existing aerator/flow straightener from screw thread tap outlet.
  • Wash hands.
  • Select appropriate Quick Connect Adaptor.
  • Insert supplied gasket and attach to outlet.
  • Insert Docking Station Quick Connect Inlet into Adaptor. Wetting the Docking Station o'ring prior to connection will ensure a good seal with the Quick Connect Adaptor.
  • An audible click will confirm engagement.
  • Ensure Docking Station front plate is positioned facing forward ready for installation of QPoint Filter Capsule.
  • Ensure o'ring seal is visible, in position and intact.
 

Disconnection

  • Press latch on Quick Connect Adaptor to release Docking Station and remove.

Disposal

  • The used Docking Station should be disposed of according to local procedures.

Pall QPoint& Filter Capsule (QJ212, QJ22)

Pictorial Installation Guide

Description

The Pall QPoint Filter Capsules are:
  • Provided sterile.
  • Disposable and may be used for up to 62 days (2 months) following initial connection.
For full details refer to the Instructions For Use.
  • Ensure the QPoint Docking Station is clean and dry to avoid inadvertent contamination.
  • Ensure o'ring seal is visible, in position and intact.
  • The Filter Capsule should be handled aseptically.
  • Wash hands and/or wear sterile gloves prior to opening filter unit packaging.
  • Whilst holding Filter Capsule securely through the unit packaging, peel back lid.
  • The unit packaging can be used to protect the device from cross contamination during installation.
  • Align marker triangle on the top face of the Filter Capsule with the marker triangle in the centre of Docking Station front panel.
  • With the triangles aligned, turn the Filter Capsule upwards onto the Docking Station while rotating a 14 turn to the right.
  • If connecting to a Docking Station with Quick Connect Inlet, the Docking Station will need to be held to avoid rotation.
  • The Filter Capsule labelling details should be aligned beneath the Docking Station front panel.
  • Remove protective unit packaging prior to use.
  • The Filter Capsule can be used for up to 62 days (2 months).
  • Suspending use following initial connection does not extend filter life.
 
The Filter Capsule should not be
  • positioned directly into a drainage point.
  • immersed in collected water.
  • placed on the floor, in a basin or bath, or in contact with contaminated surfaces or cloths.
  • touched during use.
   

Disconnection

  • Lift release button on the Docking Station to disengage Filter Capsule, and turn Filter Capsule to remove.

Disposal

  • The used Filter Capsule should be disposed of according to local procedures. All materials are suitable for plastics recycling, as per local guidelines.

Systemic Chemical Disinfection Filter Capsules

A wide variety of systemic chemical treatments are used to mitigate microbial contamination within potable water supplies to hospitals, hotels, leisure centres, schools, apartment blocks and other building complexes. Such chemical treatments may be used in conjunction with Pall Point of Use Water Filters as part of the engineering solution.

Testing has been performed to support compatibility of Pall QPoint Filter Capsules (QJ212, QJ22) with a range of common systemic chemical treatments, particularly shock treatments completed during periods of critical water system contamination.

Active chlorine at 100 parts per million (ppm)

Following exposure to active chlorine (100 ppm) for 1 hour at 22 °C (72 °F) all Filter Capsules passed. No visual abnormalities were observed.

Continuous treatment with active chlorine for up to 62 days (1 - 7 ppm)

Following exposure to up to 7 ppm active chlorine for 62 days at 22°C (72 °F) all Filter Capsules passed integrity test. No visual abnormalities were observed.

Shock treatment with pH > 12

Following exposure to pH>12 for 1 hour at 22 °C (72 °F) all Filter Capsules passed integrity test. No visual abnormalities were observed.

Shock treatment with peracetic acid (1000 ppm)

Following exposure to 1000 ppm peracetic acid for 2 hours at 60 °C (140 °F) all Filter Capsules passed integrity test. No visual abnormalities were observed.

Conclusion

Pall QPoint Filter Capsules maintain their filter integrity and are compatible with shock treatment of 100 ppm active chlorine for 1 hour, continuous active chlorine treatment of 1 - 7 ppm for 62 days, pH>12 for one hour or 1000 ppm peracetic acid for 2 hours.

Surface Disinfection Filter Capsules

  • If a filter is visibly, or known to have been contaminated, it should be replaced.
  • QPoint Filter Capsules should not come into contact with contaminated surfaces or cloths.
  • If there is a local Standard Operating Procedure in place for Filter Capsule cleansing and hygiene, which includes a new or sterile wipe/cloth for each filter outlet, then the external surfaces of the Filter Capsule (QJ212, QJ22) may be cleansed with the following common sanitising solutions containing Iso-Propyl Alcohol, Quaternary Ammonium Compounds, Benzyl Ammonium Compounds, Iodides and non-ionic surfactants. The following solution compositions (qualitative compositions) have been tested:
    • Solution A: n-propanol (29%), propionic quaternary ammoniums, polhexanide.
    • Solution B: glycolic acid, propionic quaternary ammoniums, n-propanol, non-ionic surfactants, fragrance, colour.
    • Solution C: propionic quaternary ammoniums, guanidium acetate, n-propanol, non-ionic surfactants, excipients.
    • Solution D: alkyl (60% C14, 30% C16, 5% C12, 5% C18) dimethyl benzyl ammonium chloride (2.25%), alkyl (68% C12, 32% C14) dimethylethylbenzyl ammonium chloride (2.25%), inert ingredients (95.5%).
    • Solution E: di-isobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride (0.28%), isopropanol (17.2%), inert ingredients (82.52%).
    • Solution F: povidone iodine (4%), excipients.
    • Solution G: n,n-didecyl-n-methyl-poly(oxyethyl) ammonium propionate (2.5-10%), sodium n-lauryl b-iminodipropionate (2.5-10%), ethoxylated isotridecanol (2.5-10%), monopropylene glycol (10-25%), ethane-1,2-diol,homopolymere (0-2.5%), polyhexamethylene biguanide hydrochloride (0-2.5%).
  • Periodic wiping or spraying of the housing / outlet with the solutions listed above does not adversely affect the polyester (polybutylterepthalate) housing material of the QPoint Filter Capsules. At no point should the Filter Capsule be disconnected during cleansing.

Ordering Information

 
Reorder Code Product Description Packaging
Docking Station
QDTC Tap Assembly with Quick Connect Inlet 1 unit/pkg
Filter Capsules
QJ212 Jet Outlet, 0.2 μm Filter 12 units/pkg
QJ22 Jet Outlet, 0.2 μm Filter 2 units/pkg
 
Quick Connect Adaptors
 
TAP1/2MN 1/2 inch, male thread, non-valved 1 unit
TAP1/2MV 1/2 inch, male thread, valved 1 unit
TAP1/2FN 1/2 inch, female thread, non-valved 1 unit
TAP1/2FV 1/2 inch, female thread, valved 1 unit
TAP22FN 22 mm, female thread, non-valved 1 unit
TAP22FV 22 mm, female thread, valved 1 unit
TAP24MN 24 mm, male thread, non-valved 1 unit
TAP24MV 24 mm, male thread, valved 1 unit
TAPUC Universal (15 – 29 mm), non-threaded 1 unit
TAP21MN 21 mm, male thread, non-valved 1 unit
TAP21MV 21 mm, male thread, valved 1 unit
 