White Blood Cell Isolation (Leukosorb) Medium product photo Primary L

White Blood Cell Isolation (Leukosorb) Medium

For Molecular Diagnostics

   

Our patented leukocyte removal medium is a highly wettable, fibrous matrix designed for use in procedures requiring isolation of leukocytes from whole blood samples. Whole blood samples can be applied to the top of membrane and allowed to pass through the filter matrix either by gravity or pressure. If using a pressurized filtration system, care must be employed not to lyse the red blood cells resulting in hemoglobin contamination. As the whole blood samples pass through the membrane, the leukocytes are isolated from the sample and become trapped within the fibrous matrix of the membrane. A visual representation of leukocyte removal is depicted graphically in the performance section.

The material is intended for use in assays involving analysis of the filter effluent, or direct analysis of the trapped white blood cells. It is an optimal choice for use in assays that may require the removal of white blood cells from samples to eliminate interferences. If the assay requires the analysis of the leukocytes or nucleic acid components, a variety of extraction and detection systems may be applied depending on the assay requirements. The medium can be used to trap leukocytes and extract nucleic acid samples, which can then be prepared for use in polymerase chain reaction (PCR) applications.



도움이 필요하세요?
Ask a Question
* Required

설명

For Molecular Diagnostics

   

Our patented leukocyte removal medium is a highly wettable, fibrous matrix designed for use in procedures requiring isolation of leukocytes from whole blood samples. Whole blood samples can be applied to the top of membrane and allowed to pass through the filter matrix either by gravity or pressure. If using a pressurized filtration system, care must be employed not to lyse the red blood cells resulting in hemoglobin contamination. As the whole blood samples pass through the membrane, the leukocytes are isolated from the sample and become trapped within the fibrous matrix of the membrane. A visual representation of leukocyte removal is depicted graphically in the performance section.

The material is intended for use in assays involving analysis of the filter effluent, or direct analysis of the trapped white blood cells. It is an optimal choice for use in assays that may require the removal of white blood cells from samples to eliminate interferences. If the assay requires the analysis of the leukocytes or nucleic acid components, a variety of extraction and detection systems may be applied depending on the assay requirements. The medium can be used to trap leukocytes and extract nucleic acid samples, which can then be prepared for use in polymerase chain reaction (PCR) applications.

응용분야

Applications

  • Leukocyte removal
  • Molecular diagnostics
  • DNA prep prior to PCR

Sealing

  • Mechanical

규격사양

Typical Membrane Characteristics

Base Material Pore Size (µm) Thickness (mils) Thickness (µm)
 Proprietary  8.0  14.0-22.0  355.6-558.8

Typical Performance Characteristics

Base Material Leukocyte Immobilization Leukocyte Retention Post Wash Hold-Up
 Proprietary  40-80%  70-80%  40-70 µL/cm2

Wet time: < 5.0 seconds

성능

Leukocyte Removal from Whole Blood

Removal Efficiencies of White Blood Cell Isolation Media

Sample Total Leukocytes
(x 109 L)
Percent Removal (%) Lymphocytes
(x 109 L)
Percent Removal (%)
Whole Blood Control 6.5 N/A 2.7 N/A
Membrane Filtered Sample 1 3.6 44.2 2.0 24.5
Membrane Filtered Sample 2 3.9 39.5 2.0 24.5
Membrane Filtered Sample 3 3.5 45.7 2.0 24.5
Membrane Filtered Sample 4 3.7 42.6 2.0 24.5
Sample Mid-Range Leukocytes
(x 109 L)
Percent Removal (%) Granulocytes
(x 109 L)
Percent Removal (%)
Whole Blood Control 2.0 N/A 1.8 N/A
Membrane Filtered Sample 1 0.9 56.1 0.7 42.9
Membrane Filtered Sample 2 1.1 46.3 0.8 54.3
Membrane Filtered Sample 3 1.0 51.2 0.5 40.0
Membrane Filtered Sample 4 1.1 46.3 0.6 52.2

종류

Membrane Filters

사용

Filtration, Separation, Removal, Diagnostics

주문정보

Custom roll, sheet, and disc sizes available upon request. Please contact your local sales representative for additional information.

Part Number Description Pkg
 BSP0669  White Blood Cell Isolation Media, 8" x 10" sheet  5/pkg

적용분야

Diagnostics - Plasma Seperation, Diagnostics - Sample Preparation

영역

OEM Diagnostics Components, OEM Diagnostics Components

Reviews