Acro® 50 Vent Filters
* Required
Need Help?
Acro® 50 vent filters, 0.2 µm sterilizing grade, 19.6 cm² EFA, ½-¼ in. stepped hose barb connections, box of 72
Product ID: 6074270L
Unit of Measure
1/PAC
1/PAC
Min Order Qty
1
1
Membrane Material
PTFE
Industry-Specific Certification
Pharmaceutical Grade
Effective Filtration Area (Metric)
19.6 cm²
Format
Acro 50
Inlet Connection
¼-½ in (6-13 mm) hose barb
Outlet Connection
¼-½ in (6-13 mm) hose barb
Description
Pall Acro 50 disposable filters are 0.2 micron rated and contain a Pall hydrophobic PTFE membrane and are ideal for the sterile venting of small containers and as an in-line gas filter. The large effective filter area provides high flow rates and filters can be autoclaved several times ensuring low costs.
Features
- PTFE membrane provides high flow rates and resists ‘wetting’ out by water vapor
- Light compact unit suitable for small vessels and flexible tubing
- Can be easily integrity tested by water breakthrough test
- Each filter labelled with an individual serial number
High Quality Standards
- Validated with Brevundimonas diminuta (ATCC 19146) in liquid at a challenge level of >107/cm2 filter area
- 100% integrity tested
- Manufactured in a clean environment under controlled conditions
- Materials of construction tested for biological compatibility
- Comprehensive validation guide available
Specifications
1
2
Materials of Construction
Membrane | PTFE on a polypropylene support |
Housing | Polypropylene |
Operating Parameters1
Max operating pressure | 4.1 bar at up to 25 °C (77°F) |
Max operating temperature | 40 °C (104 ºF) |
1 In compatible fluids which do not soften, swell or adversely affect the filter or its materials of construction.
Autoclave Sterilization2
Autoclave: 5 x 0.5-hour cycles at 125°C (257°F)
2 The autoclave life was determined by testing under controlled laboratory conditions up to the time indicated. Actual operating conditions may affect the filter capsule's resistance to autoclave sterilization and service. Filter capsules should be qualified for each process. These filter capsules should not be exposed to gamma irradiation.
Warning: This product must not be sterilized in-situ by passing steam through under pressure. Integrity should be verified after each autoclave cycle before use.
Connections
Inlet and Outlet: ¼ in. - ½ in. hose barb
Typical Effective Filtration Area
19.5 cm2
Removal Rating
0.2 micron
Typical Air Flow Rates
>8 standard liters /minute at 0.2 bard (3 psid) differential pressure
Nominal Dimensions
Application
Filters - Gas
Product Summary
0.2 micron-rated gas filters, ideal for the sterile venting of small containers or as an in-line gas filter