Fluorodyne® II DBL - Junior Filters (MCY) product photo Primary L

Fluorodyne® II DBL - Junior Filters (MCY)

These 0.45 micron-rated filters are predominantly used in pre-filtration applications or viscous fluid filtrations. Validated removal of Serratia marcescens (ATCC 14756)
with a typical titer reduction (TR) of > 106 ensures that they are also suitable for bioburden control. A serial layer (0.65/0.45 micron) membrane construction allows high throughputs and high flow rates compared with other 0.45 micron-rated filters.

Junior filter cartridges (MCY style) are single-use filter cartridges, for use with in-line stainless steel filter housings, and feature double O-ring piston seals comparable to ‘AB-style’ cartridges. They are designed for processing of volumes of material (30 L to 500 L) in a small-to-mid-size GMP manufacturing environment.


Need Help?
Ask a Question
* Required

Description

These 0.45 micron-rated filters are predominantly used in pre-filtration applications or viscous fluid filtrations. Validated removal of Serratia marcescens (ATCC 14756)
with a typical titer reduction (TR) of > 106 ensures that they are also suitable for bioburden control. A serial layer (0.65/0.45 micron) membrane construction allows high throughputs and high flow rates compared with other 0.45 micron-rated filters.

Junior filter cartridges (MCY style) are single-use filter cartridges, for use with in-line stainless steel filter housings, and feature double O-ring piston seals comparable to ‘AB-style’ cartridges. They are designed for processing of volumes of material (30 L to 500 L) in a small-to-mid-size GMP manufacturing environment.

Specifications

Materials Of Construction

 
Filter Membrane Hydrophilic modified PVDF
Support/Drainage Polypropylene
End Cap, Core and Cage Polypropylene
Sealing Technology Thermal bonding without adhesives
O-rings Silicone elastomer
 

Operating Parameters¹

 
Maximum Operating Temperature  80 °C 
Maximum Differential Pressure (Forward Direction) 5.3 bar (77 psi) at 40 ºC
3.4 bar (50 psi) at 80 ºC
 
¹ In compatible fluids which do not soften, swell or adversely affect the filter or its materials of construction


Sterilization²

 
Autoclave and In Situ steam 30 x 60 minutes at 125 °C
10 x 60 minutes at 140 °C
 

Typical Extractables in Water at 20 °C³

 
< 5 mg per Junior cartridge
 
³ Tested on elements without pre-flushing

Nominal Dimensions

 
Length 120 mm (4.7 in.) 
 

Nominal Effective Filter Area (EFA)

 
1500 cm² (1.6 ft²) 
 

Type

Liquid Filters

Use

Bioburden Filtration

Ordering Information

 
Item Part Number
Junior filter cartridge (MCY style)
Single open-ended cartridge, double external o-rings (silicone elastomer) and locating lugs, 0.45 micron bioburden control
MCY4440DBLPH4
 

Product Summary

MCY style junior cartridges are ideal single-use productions for prefiltration and suitable for bioburden control. Fluorodyne® II DBL Membrane in Junior Filter Cartridges provide reliable, sterile filtration for viscous fluids. Selective membrane construction allows for high throughputs and flow rates. MCY junior cartridges are used in stainless steel housings for small to medium sized operations.

Reviews