Fluorodyne® II DJL Membrane in Junior Filter Cartridges (MCY Style) product photo Primary L

Fluorodyne® II DJL Membrane in Junior Filter Cartridges (MCY Style)

These 0.1 micron-rated filters with serial layer (0.2/0.1 micron) membrane construction assure high flow rates compared with other 0.1 micron (and even some 0.2 micron) -rated filters. Fluorodyne II grade DJL filters are validated for removal of Acholeplasma laidlawii mycoplasma (ATCC 28206) with a typical TR >1011 and retention of Brevundimonas diminuta (ATCC 19146) at 107 CFU/cm² effective filter area (EFA). This allows for enhanced sterilization assurance as well as efficient mycoplasma control.

Junior filter cartridges (MCY style) are single-use filter cartridges, for use with in-line stainless steel filter housings, and feature double O-ring piston seals comparable to ‘AB-style’ cartridges. They are designed for processing of volumes of material (30 L to 500 L) in a small-to-mid-size GMP manufacturing environment.



* Required
Need Help?
Description

These 0.1 micron-rated filters with serial layer (0.2/0.1 micron) membrane construction assure high flow rates compared with other 0.1 micron (and even some 0.2 micron) -rated filters. Fluorodyne II grade DJL filters are validated for removal of Acholeplasma laidlawii mycoplasma (ATCC 28206) with a typical TR >1011 and retention of Brevundimonas diminuta (ATCC 19146) at 107 CFU/cm² effective filter area (EFA). This allows for enhanced sterilization assurance as well as efficient mycoplasma control.

Junior filter cartridges (MCY style) are single-use filter cartridges, for use with in-line stainless steel filter housings, and feature double O-ring piston seals comparable to ‘AB-style’ cartridges. They are designed for processing of volumes of material (30 L to 500 L) in a small-to-mid-size GMP manufacturing environment.

Specifications

Materials Of Construction

 
Filter Membrane Hydrophilic modified PVDF
Support/Drainage Polypropylene
End Cap, Core and Cage Polypropylene
Sealing Technology Thermal bonding without adhesives
O-rings Silicone elastomer
 

Operating Parameters¹

 
Maximum Operating Temperature  80 °C 
Maximum Differential Pressure (Forward Direction) 5.3 bar (77 psi) at 40 ºC
3.4 bar (50 psi) at 80 ºC
 
¹ In compatible fluids which do not soften, swell or adversely affect the filter or its materials of construction.


Sterilization

 
Autoclave and In Situ steam 30 x 60 minutes at 125 °C
10 x 60 minutes at 140 °C²
 
² MCY style only

Typical Extractables in Water at 20 °C³

 
< 5 mg per Junior cartridge
 
³ Tested on elements without pre-flushing

Nominal Effective Filter Area (EFA)

 
MCY1110  400 cm² (0.4 ft²) 
MCY4440 1500 cm² (1.6 ft²)
 

Nominal Dimensions

 
  MCY1110 MCY4440
Length 48 mm (1.9 in.) 120 mm (4.7 in.)
 

Ordering Information
 
Item Part Number
Junior filter cartridge (MCY style)
Single open-ended cartridge, double external o-rings (silicone elastomer) and locating lugs, 0.1 micron sterilizing grade
MCY1110DJLPH4
Junior filter cartridge (MCY style)
Single open-ended cartridge, double external o-rings (silicone elastomer) and locating lugs, 0.1 micron sterilizing grade
MCY4440DJLPH4
 
Application
Filters - Liquid - Mycoplasma Removal (0.1 micron)