Supor® EKV Membrane Filters product photo Primary L

Supor® EKV Membrane Filters

Cost-Effective Filtration of Buffers, Tissue Culture Media, and Other Biological Fluids

  • Disposable – Reduces labor costs associated with assembling, cleaning, and testing stainless steel filter holders.
  • Flexible – Encapsulated format for ease of use.
  • Extended life – Built-in MachV asymmetric prefilter layer for maximum flow and throughput performance.
  • Simplified scale-up – Membrane and housing materials are identical to those used in larger-volume capsules and cartridges from Pall Life Sciences.
  • Low adsorption and high chemical compatibility – Featuring hydrophilic polyethersulfone membrane for high compatibility over the entire pH range and low protein binding.
  • Presterilized – Using gamma irradiation, products are presterilized, non-pyrogenic, and non-cytotoxic (with the exception of 47 mm disc filters).
  • Inherently hydrophilic – Provides a reliable integrity test. 
  • Tested – Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>. Assessed during manufacture to ensure product integrity.
“Pall Laboratory products are designed for laboratory applications only. This product is not approved for use in medical, clinical, surgical or other patient applications. If you require further assistance in product selection for your chosen application, please contact Labcustomersupport@pall.com or call 1-800-521-1520."


* Required
Need Help?
Download Datasheet

AcroPak 1500 Capsules With Supor EKV Membrane

List Price
$ 200.00
List Price
$ 200.00

AcroPak 20 Filters With Supor EKV Membrane

List Price
$ 55.00
List Price
$ 62.00

AcroPak 200 Capsules With Supor EKV Membrane

List Price
$ 222.00
List Price
$ 222.00

AcroPak 400 Capsules With Supor EKV Membrane

List Price
$ 106.00
List Price
$ 106.00

AcroPak 800 Capsules With Supor EKV Membrane

List Price
$ 160.00
List Price
$ 160.00

Acrodisc® Syringe Filters, 25 mm

List Price
$ 170.00

Membrane Stack Disc Filters

List Price
$ 35.97
Description

Cost-Effective Filtration of Buffers, Tissue Culture Media, and Other Biological Fluids

  • Disposable – Reduces labor costs associated with assembling, cleaning, and testing stainless steel filter holders.
  • Flexible – Encapsulated format for ease of use.
  • Extended life – Built-in MachV asymmetric prefilter layer for maximum flow and throughput performance.
  • Simplified scale-up – Membrane and housing materials are identical to those used in larger-volume capsules and cartridges from Pall Life Sciences.
  • Low adsorption and high chemical compatibility – Featuring hydrophilic polyethersulfone membrane for high compatibility over the entire pH range and low protein binding.
  • Presterilized – Using gamma irradiation, products are presterilized, non-pyrogenic, and non-cytotoxic (with the exception of 47 mm disc filters).
  • Inherently hydrophilic – Provides a reliable integrity test. 
  • Tested – Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>. Assessed during manufacture to ensure product integrity.
“Pall Laboratory products are designed for laboratory applications only. This product is not approved for use in medical, clinical, surgical or other patient applications. If you require further assistance in product selection for your chosen application, please contact Labcustomersupport@pall.com or call 1-800-521-1520."
Applications

47 mm Disc Filters

  • High throughput
  • Low protein binding
  • Fast flow rates

Applications

Suitable for sterile filtration of a wide range of fluids, including:
  • Buffers  
  • Biological fluids
  • Cell culture media
  • Ophthalmic products

AcroPak 20 Filters

AcroPak 20 Filters with Supor EKV Membrane Filters
  • Designed to add convenience to small- and mediumvolume filtration
  • Generates twice the effective filtration area and double the throughput/flow rate of conventional metal 47 mm reusable filter holders
  • Disposable for reduced labor
  • Significantly reduces the cost of repeat testing when scaling up your filtration system

Applications

  • Ideal for filtering aqueous solutions, cell culture media, and serum
  • Small- to medium-volume sterile filtration, typically up to 1 liter
  • Suitable for use with fluids containing dilute proteins, preservatives, or other critical components
  • Sterile filtration of media and buffers

25 mm Acrodisc® Syringe Filters

25 mm Acrodisc Syringe Filters with Supor Membrane Filter
  • Universal life science membrane with low protein binding to minimize sample loss
  • Superior flow rates and higher throughputs than similar devices
  • Sterilized by gamma irradiation to eliminate potential contamination by EtO residuals

Applications

  • Small-volume applications (such as liquid sterilization and determination of product compatibility and recovery)
  • Ideal for typical process volume up to 100 mL

AcroPak 200 Capsules

  • High filtration area and compact size are ideal for upscale trials
  • Minimal hold-up volume
  • Good wettability for reliable integrity test
  • United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
  • Sealed membrane and housing uses fusion technology to eliminate potential extractables
  • Gamma sterilized, non-pyrogenic, and provided with a removable filling bell (except sanitary flange option)
  • Hydrophilic polyethersulfone membrane for low adsorption and wide chemical compatibility
  • Very high flow rates and consistently higher total solution throughputs due to superior porosity over other membranes

Applications

  • Designed to quickly process difficult-to-filter solutions within 5-21 liters, such as serum, serum-supplemented culture media, and ascites fluid
  • Ideal in situations where rapid filtration or short processing times are essential

AcroPak 400, 800 and 1500 Capsules

  • Encapsulated format for higher flexibility, minimized cleaning, and low installation costs
  • Minimal hold-up volume
  • Hydrophilic polyethersulfone membrane for low adsorption and wide chemical compatibility
  • Good wettability for a reliable integrity test
  • Patented Ultipleat technology for high area and good flow rates
  • Built-in MachV asymmetric prefilter layer for long-life and low filtration costs

Applications

Suitable for sterile filtration of a wide range of fluids, including:
  • Buffers
  • Biological fluids
  • Cell culture media
  • Ophthalmic products
Specifications
 
47 mm Disc Filters with Supor EKV Membrane


47 mm Disc Filters with Supor EKV Membrane
Filter Media Supor EKV membrane (hydrophilic polyethersulfone)
Pore Size 0.2 μm
Effective Filtration Area 9.6 cm2 (1.4 in.2) (dependent on holder
 

 
25 mm Acrodisc Syringe Filterswith Supor EKV Membrane


25 mm Acrodisc Syringe Filterswith Supor EKV Membrane
Filter Media Supor EKV membrane (hydrophilic polyethersulfone)
Housing, Vent Plug, and Support Material Polypropylene
Pore Size 0.2 μm
Effective Filtration Area 2.8 cm2
Dimensions Length: 21 mm (0.8 in.)
Diameter: 29 mm (1.2 in.)
Inlet/Outlet Connections Female luer lock inlet, male slip luer outlet
Typical Hold-up Volume (with Air Purge) < 100 μL
Maximum Operating Temperature* 60 ºC (140 ºF) at 2.1 bar (210 kPa, 30 psi)
Maximum Operating Pressure* 5.4 bar (540 kPa, 80 psi) at ambient temperature
Recommended Integrity Test Minimum Bubble Point (Water) > 3.32 bar (332 kPa, 48 psi)
Bacterial Retention Lot samples retain a minimum of 107 cfu/cm2 of B. diminuta per modified ASTM F838, current revision
Endotoxin Level < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Sterilization Sterilized by gamma irradiation and individually packaged
 

 
AcroPak 20 Filters with Supor EKV Membrane


AcroPak 20 Filters with Supor EKV Membrane
Filter Media Supor EKV membrane (hydrophilic polyethersulfone)
Housing, Vent Plug, and Support Material Polypropylene
Filling Bell Polycarbonate
Sealing Technology Thermal bonding without adhesives
Pore Size 0.2 μm
Effective Filtration Area 20 cm2
Dimensions Length: 83 mm (3.3 in.)
Diameter: 67 mm (2.7 in.)
Inlet/Outlet Connections Stepped hose barbs, 6.4-12.7 mm (1/4-1/2 in.) diameter with female slip luer ID in the hose barb
Typical Hold-up Volume (with Air Purge) < 2.5 mL
Maximum Operating Temperature* 60 ºC (140 ºF) at 1.0 bar (100 kPa, 15 psi) at 21-24 ºC (70-75 ºF)
Maximum Operating Pressure* 4.1 bar (410 kPa, 60 psi)
Recommended Integrity Test Minimum Bubble Point (Water) > 3.32 bar (332 kPa, 48 psi)
Bacterial Retention Lot samples retain a minimum of 107 cfu/cm2 of B. diminutaper modified ASTM F838, current revision
Endotoxin Level < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Sterilization Sterilized by gamma irradiation or sold non-sterile; if desired, autoclave once at 125 ºC (257 ºF) for 60 minutes; water wet capsule prior to autoclaving
 

 
AcroPak 200 Capsules with Supor EKV Membrane


AcroPak 200 Capsules with Supor EKV Membrane
Filter Media Supor EKV membrane (hydrophilic polyethersulfone)
Housing, Vent Plug, and Support Material Polypropylene
Filling Bell Polycarbonate (No filling bell on sanitary flange option)
Sealing Technology Thermal bonding without adhesives
Pore Size 0.2 μm
Effective Filtration Area 220 cm2
Dimensions Housing Length
   12094: 10.5 cm (4.1 in.)
   12095: 7.3 cm (2.9 in.)
Housing Diameter
   5.3 cm (2.1 in.) without vent
   6.7 cm (2.6 in.) with vent
Inlet/Outlet Connections 12460: 25-38 mm (1-1.5 in.) sanitary flange inlet/outlet connection
12461: 13 mm (1/2 in.) hose barb inlet/outlet connection
Typical Hold-up Volume (with Air Purge) < 6 mL
Maximum Operating Temperature* 60 ºC (140 ºF) at 2.1 bar
Maximum Operating Pressure* 4.1 bar (410 kPa, 60 psi) at ambient temperature
Recommended Integrity Test Minimum Bubble Point (Water) > 3.32 bar (332 kPa, 48 psi) (210 kPa, 30 psi)
Typical Water Flow Rate 350 mL/min/0.1 bar (241 mL/min/psi)
Bacterial Retention Lot samples retain a minimum of 107 cfu/cm2 of B. diminutaper modified ASTM F838, current revision
Endotoxin Level < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Sterilization Sterilized by gamma irradiation and individually packaged; if desired, autoclave once prior to use at 131 ºC (268 ºF) for 30 minutes; water wet capsule prior to autoclaving
 

 
AcroPak 400 Capsules with Supor EKV Membrane


AcroPak 400 Capsules with Supor EKV Membrane
Filter Media Supor EKV membrane (hydrophilic polyethersulfone)
Housing, Vent Plug, and Support Material Polypropylene
Filling Bell Polycarbonate (No filling bell on sanitary flange option)
Sealing Technology Thermal bonding without adhesives
Pore Size 0.2 μm
Effective Filtration Area 375 cm2
Dimensions Housing Length
   12460: 11.7 cm (4.6 in.)
   12461: 15.7 cm (6.2 in.)
Housing Diameter (including valves)
    9.4 cm (3.7 in.)
Inlet/Outlet Connections 12460: 25-38 mm (1-1.5 in.) sanitary flange inlet/outlet connection
12461: 13 mm (1/2 in.) hose barb inlet/outlet connection
Maximum Operating Temperature* 40 ºC (104 ºF) at 4.0 bar (400 kPa, 58 psi)
Maximum Operating Pressure* 5.2 bar (520 kPa, 75 psi) at 20 ºC (68 ºF)
Recommended Integrity Test Minimum Bubble Point (Water) > 3.32 bar (332 kPa, 48 psi)
Typical Water Flow Rate 700 mL/min/0.1 bar (483 mL/min/psi)
Bacterial Retention Lot samples retain a minimum of 107 cfu/cm2 of B. diminutaper modified ASTM F838, current revision
Endotoxin Level < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Sterilization Sterilized by gamma irradiation and individually packaged
 

 
AcroPak 800 Capsules with Supor EKV Membrane


AcroPak 800 Capsules with Supor EKV Membrane
Filter Media Supor EKV membrane (hydrophilic polyethersulfone)
Housing, Vent Plug, and Support Material Polypropylene
Filling Bell Polycarbonate (No filling bell on sanitary flange option)
Sealing Technology Thermal bonding without adhesives
Pore Size 0.2 μm
Effective Filtration Area 750 cm2
Dimensions Housing Length
   12463: 15.7 cm (6.2 in.)
   12464: 19.7 cm (7.7 in.)
Housing Diameter (including valves)
   9.4 cm (3.7 in.)
Inlet/Outlet Connections

12463: 25-38 mm (1-1.5 in.) sanitary flange inlet/outlet connection
12464: 13 mm (1/2 in.) hose barb inlet/outlet connection

Maximum Operating Temperature* 40 ºC (104 ºF) at 4.0 bar (400 kPa, 58 psi)
Maximum Operating Pressure* 5.2 bar (520 kPa, 75 psi) at 20 ºC (68 ºF)
Recommended Integrity Test Minimum Bubble Point (Water) > 3.32 bar (332 kPa, 48 psi)
Typical Water Flow Rate 1.5 L/min/0.1 bar (1.0 L/min/psi)
Bacterial Retention Lot samples retain a minimum of 107 cfu/cm2 of B. diminutaper modified ASTM F838, current revision
Endotoxin Level < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Sterilization Sterilized by gamma irradiation and individually packaged
 

 
AcroPak 1500 Capsules with Supor EKV Membrane


AcroPak 1500 Capsules with Supor EKV Membrane
Filter Media Supor EKV membrane (hydrophilic polyethersulfone)
Housing, Vent Plug, and Support Material Polypropylene
Filling Bell Polycarbonate (No filling bell on sanitary flange option)
Sealing Technology Thermal bonding without adhesives
Pore Size 0.2 μm
Effective Filtration Area 1500 cm2
Dimensions Housing Length
   1246: 17.4 cm (6.8 in.)
   12467: 21.0 cm (8.3 in.)
Housing Diameter (including valves)
   10.5 cm (4.1 in.)
Inlet/Outlet Connections 12466: 25-38 mm (1-1.5 in.) sanitary flange inlet/outlet connection
12467: 13 mm (1/2 in.) hose barb inlet/outlet connection
Maximum Operating Temperature* 40 ºC (104 ºF) at 4.0 bar (400 kPa, 58 psi)
Maximum Operating Pressure* 5.2 bar (520 kPa, 75 psi) at 20 ºC (68 ºF)
Recommended Integrity Test Minimum Bubble Point (Water) > 3.32 bar (332 kPa, 48 psi)
Typical Water Flow Rate 3.2 L/min/0.1 bar (2.2 L/min/psi)
Bacterial Retention Lot samples retain a minimum of 107 cfu/cm2 of B. diminutaper modified ASTM F838, current revision
Endotoxin Level < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Sterilization Sterilized by gamma irradiation and individually packaged
 

*In compatible fluids which do not soften, swell or adversely affect the filter or its materials of construction.
Type
Membrane Disc
Use
Microfiltration
Application
Sterile Filtration
Segment