Vacuum Protection Filter product photo Primary L

Vacuum Protection Filter

Superior air flow rates and low back pressure maximizes application performance.

  • Low system pressure resistance - maximizes pump performance
  • Effective barrier - protects vacuum pumps from biohazardous contamination
  • Filter delivers viral and bacterial protection at 15 psi
  • Custom colored printing option for product and/or company identification
  • Assists in meeting 29CFR1910.1030 OSHA Bloodborne Pathogen Standard by serving as an "Engineering Control" to reduce employee exposure
  • Meets bacterial filter requirements suggested by the Center for Devices and Radiological Health Guidance Development for Powered Suction Pump 510(k)s


* Required
Need Help?
Description

Superior air flow rates and low back pressure maximizes application performance.

  • Low system pressure resistance - maximizes pump performance
  • Effective barrier - protects vacuum pumps from biohazardous contamination
  • Filter delivers viral and bacterial protection at 15 psi
  • Custom colored printing option for product and/or company identification
  • Assists in meeting 29CFR1910.1030 OSHA Bloodborne Pathogen Standard by serving as an "Engineering Control" to reduce employee exposure
  • Meets bacterial filter requirements suggested by the Center for Devices and Radiological Health Guidance Development for Powered Suction Pump 510(k)s
Applications
  • Vacuum protection
  • General hydrophobic barrier
  • Small to medium volume vent
Specifications

Bacterial/Viral Filtration Efficiency
Bacterial removal efficiency 99.999% with test organisms Brevundimonas diminuta.
Test organism: Serratia marcescens (0.45 µm) Brevundimonas diminuta (0.30 µm)

Viral Filtration Efficiency
Viral removal efficiency 99.999% with MS-2 Bacteriophage.
Test organism: MS-2 Bacteriophage

HEPA
Exceeds filtration efficiency requirements for HEPA rating (data available upon request)

Effective Filtration Area
15 cm2

Materials of Construction
Media: PTFE with a prefilter support
Housing: Acrylic

Inlet/Outlet Connections
1/4 in. (6.4 mm) HB
Orientation: Blue side toward vacuum source

Typical Air Flow Rate
30 L/min @ 5 psi

Minimum Water Breakthrough
750 mmHg (15 psi)

Pyrogenicity
< 0.25 EU/mL using the LAL test method

Sterilization Compatibility
EtO

Biological Safety
Materials of construction pass USP Biological Reactivity Tests, in Vivo <88>

Type
In-line Filters
Use
Filtration
Ordering Information
Product No. Description
8001052 Vacuum Protection Filter, bulk, non-sterile
Application
Medical Venting and Gas Filtration