IV-3 Filter product photo Primary L

IV-3 Filter

The Original Self-priming Filter Requires No Manipulation to Provide Position Insensitive Air Removal and Fluid Delivery.

  • 0.2 µm Supor® membrane is retentive of B. diminuta and meets USP 25/NF20 requirements for a sterilizing grade filter per modified ASTM F838-83 test methods
  • Maximizes delivery of critical drugs with proven low drug binding Supor membrane
  • Contains no exogenous surfactants, protecting patient from delivery of unwanted extractables
  • Colored housings and pad printing available for ease of final product identification
  • Materials of construction contain no natural rubber latex, or latex derivatives, PVC, or hydroxyvinyl chloride


* Required
Need Help?
Download Datasheet
Description

The Original Self-priming Filter Requires No Manipulation to Provide Position Insensitive Air Removal and Fluid Delivery.

  • 0.2 µm Supor® membrane is retentive of B. diminuta and meets USP 25/NF20 requirements for a sterilizing grade filter per modified ASTM F838-83 test methods
  • Maximizes delivery of critical drugs with proven low drug binding Supor membrane
  • Contains no exogenous surfactants, protecting patient from delivery of unwanted extractables
  • Colored housings and pad printing available for ease of final product identification
  • Materials of construction contain no natural rubber latex, or latex derivatives, PVC, or hydroxyvinyl chloride
Applications
  • IV solutions
  • TPN and lipid solutions
  • Drug therapy
  • Apheresis solutions
Specifications

Biological Safety
Materials of construction pass USP Biological Reactivity Tests, In Vivo <88>

Dimensions
33 x 66 mm (1.3 X 2.6 in.)

Effective Filtration Area
10 cm2

Inlet/Outlet Connections
Standard Connectors:
Outer Diameter: 3.43 mm (0.135 in.)
Inner Diameter: 1.40 mm (0.055 in.)

Materials of Construction
Filter Media: Supor® membrane
Housing: Modified acrylic
Vent: 0.02 µm supported PTFE

Maximum Operating Pressure
3.1 bar (45 psi)

Maximum Operating Temperature
55 °C (131 °F)

Minimum Water Bubble Point
0.2 µm: ≥ 3.1 bar (45 psi)
0.8 µm: ≥ 0.7 bar (10 psi)
1.2 µm: ≥ 0.3 bar (5 psi)

Minimum Water Flow Rate
0.2 µm: ≥ 20 mL/min at 91 cm (36 in.) head pressure
0.8 µm: ≥ 50 mL/min at 91 cm (36 in.) head pressure
1.2 µm: ≥ 100 mL/min at 91 cm (36 in.) head pressure

Pore Size
0.2, 0.8, 1.2 µm
Customized products available

Priming Volume
2.4 mL

Pyrogenicity
< 0.25 EU/mL using the LAL test method

Sterilization Compatibility
EtO, gamma irradiation (tested to 50 kGy)

Type
In-line Filters
Use
Filtration, Delivery
Ordering Information
Part Number Pore Size Description
6494420 0.2 µm Supor membrane, clear housing, SIV connectors
6584420 0.8 µm Supor membrane, clear housing, SIV connectors
6974420 +0.2 µm Charged Supor membrane, clear/green housing, SIV connectors
6564420 1.2 µm Supor membrane, clear/blue housing, SIV connectors
6102042 Clip Clip for IV-3 filters
Application
Medication and IV Fluid Delivery Filtration