Supracap™ 50 Depth Filter Capsules product photo Primary L

Supracap™ 50 Depth Filter Capsules

Small Volume and Scalable

Supracap 50 capsules are designed for developing and optimizing a process during scale-up and scale-down studies. They can be used to quickly and accurately determine which series and grade of depth filter media will provide the best performance as well as the necessary filtration area required to meet process volume.

Supracap 50 capsules are readily scalable to Pall Supracap 100 capsules, Supradisc™ I, and Supradisc II depth filter modules. Supracap 50 capsules are also available with Seitz® HP sheet media (double-layer) that scales to Pall Supradisc HP filter modules.  Supracap 50 capsules are available with a wide selection of Seitz depth filter media for optimal grade selection.

High Quality Standards

  • Batch tested in order to meet all quality requirements
  • Manufactured under a Quality Management System certified to ISO 9000
  • All plastic components used in construction meet the specifications for Biological Reactivity Tests in vivo for Class VI Plastics (121 °C) as described in the current United States Pharmacopoeia (USP).

Fields of Application

 

Pall UpScaleSM Program

From drug discovery and basic research, through process development and production, Pall Corporation is the single source for all your filtration and separation needs. Our UpScale program provides you with the scalable products and support you need to bring new products to market faster.

Pall Activated Carbon Filters

Supracap 50 capsules are also available as part of the scalable Pall activated carbon filter range, incorporating Seitz® AKS filter media in 9 different grades of carbon.

Further Information

Further information is available on Pall's range of Filter Modules and Capsules with Seitz Depth Filter Media and also the Pall Supracap Depth Filter Capsules range of products.



Contact Us
* Required

Order Products

Documents

Datasheets

Download: English

Description

Small Volume and Scalable

Supracap 50 capsules are designed for developing and optimizing a process during scale-up and scale-down studies. They can be used to quickly and accurately determine which series and grade of depth filter media will provide the best performance as well as the necessary filtration area required to meet process volume.

Supracap 50 capsules are readily scalable to Pall Supracap 100 capsules, Supradisc™ I, and Supradisc II depth filter modules. Supracap 50 capsules are also available with Seitz® HP sheet media (double-layer) that scales to Pall Supradisc HP filter modules.  Supracap 50 capsules are available with a wide selection of Seitz depth filter media for optimal grade selection.

High Quality Standards

  • Batch tested in order to meet all quality requirements
  • Manufactured under a Quality Management System certified to ISO 9000
  • All plastic components used in construction meet the specifications for Biological Reactivity Tests in vivo for Class VI Plastics (121 °C) as described in the current United States Pharmacopoeia (USP).

Fields of Application

 

Pall UpScaleSM Program

From drug discovery and basic research, through process development and production, Pall Corporation is the single source for all your filtration and separation needs. Our UpScale program provides you with the scalable products and support you need to bring new products to market faster.

Pall Activated Carbon Filters

Supracap 50 capsules are also available as part of the scalable Pall activated carbon filter range, incorporating Seitz® AKS filter media in 9 different grades of carbon.

Further Information

Further information is available on Pall's range of Filter Modules and Capsules with Seitz Depth Filter Media and also the Pall Supracap Depth Filter Capsules range of products.

Specifications

Typical1 Values for Ions after Flushing with 50 L/m2 WFI (Supracap 50 HP with 100 L/m2 WFI)

 
Ca Mg Fe Cu Al Ni
<0.5 ppm  < 0.1 ppm  < 0.1 ppm  < 0.1 ppm  < 50 ppb  < 10 ppb 
 
1 The tests were carried out according to the methods of the Technical/Analytical Work Group within the European Depth Filter Association, or in accordance with in-house test methods.

Materials of Constructions

 
 Media  Cellulose base 2
 Capsule  Polypropylene
 Vent  Polypropylene
 Sealing technology  Thermal bonding
 
2 See ordering information

Operating Parameters 3

 
Maximum operating pressure and temperature  3 bar (44 psi) at 40 oC
Maximum differential pressure  1.5 bar (22 psi)
 
3 In compatible fluis which do not soften, swell or adversely affect the filter or its materials of construction

Sterilization

 
Autoclaving at 125 o  1 cycle x 30 minutes
 

Typical Flow Rates

Flow dp HP series






Technical Specification

 
Code4 Depth Filter Type Typical Water Permeability in L/min/m2 @ Δp 1 bar (14.5 psid) Nominal Retention Rating in µm Ash Content in % Endotoxin Level in EU/mL before Rinsing
PDAX PDAX 15 0.05 – 0.3 60 < 0.06
PDD1 PDD1  25 0.1-0.85 52 <0.06
PDE1 PDE1 27 0.1 – 3.0 52 < 0.06
PDE2 PDE2 35 0.2-3.5 47 <0.06 
PDH4 PDH4 93 0.5-15.0 45 <0.06
PDK5 PDK5 151 1.5-20.0 45 <0.06
PDK7 PDK7 463 3 - 20 45 <0.06
PDP8 PDP8 1010 6 - 30 39 <0.06
PDK11 PDK11 173 2 - 20 49 <0.06
PEKX EKXP 25 0.05 – 0.2 62 < 0.06
PEKS EKSP 29 0.1-0.3 58 <0.06
PEKM EKMP 41 0.2-0.5 48 <0.06
PEK1 SUPRA EK1P 64 0.2-0.5 47 <0.06
P050 KS 50P 93 0.4-0.8 46 <0.06
P080 SUPRA 80P 159 1.0-3.0 49 <0.06
P100 K100P 149 1.0-3.0 45 <0.06
P200 K200P 217 3.0-6.0 43 <0.06
P250 K250P 535 4.0-9.0 44 <0.06
P700 K700P 935 6.0-15.0 45 <0.06
P900 K900P 1980 8.0-20.0 45 <0.06
B010 BIO 10 30 0.2-0.4 <1 <0.06
B020 BIO 20 75 0.4-1.0 <1 <0.06
V100 V100 P 173 2-4 53 <0.06
 

NA - Not Applicable as endotoxin content is not a measured criteria for these sheets
4 Other grades available on request

Dual-Layer Filter Media Grade Combinations

Media Grade Codes for Stax™, Supracap™ and Supradisc™ Modules

A dual layer media grade is a combination of two single layer media grades. The media grade code starts with P (Pharmaceutical Grade) followed by D (Dual layer) and the appropriate dual media combination ID. For generating a dual media combination ID, the appropriate top sheet grade letter (columns) is combined with the bottom sheet grade number or “x” (lines).


Nominal Dimensions

 
 A Supracap 50 single layer
Supracap 50 HP
42 mm (1.65 in.)
52 mm (2.05 in.)
 B 64 mm (2.52 in.)  
 

Typical Filtration Area

 
22 cm2 (3.41 in.2)
 

Connections

 
Venting, inlet and outlet Luer-lock
 

Type

Sheet Depth Filter Modules

Use

Depth Filtration

Product Summary

Supracap 50 depth filter capsules are designed for developing and optimizing a process during scale-up and scale-down studies. They can be used to quickly and accurately determine which series and grade of depth filter media will provide the best performance as well as the necessary filtration area required to meet process volume

Reviews