Product image is a representative example for illustration purposes. Actual product may vary.
Acro® 50 Vent Devices with Emflon® II Membrane - Hose barb, no vent, 0.2 µm 100% integrity tested during manufacturing (100/pkg)
Product ID: A50V002P2NV Need this product with different specifications
Connections
Stepped hose barb
EFA (Effective Filtration Area)
20 cm²
Media
Emflon II (PVDF)
Pore Size (Removal Rating)
0.2
Quantity
100 pkg
Size (Metric mm)
50 mm
Sterile
Non-Sterile
Description
Acro 50 In Line Vent Filter with Emflon II Membrane, Hose barb, no vent, 0.2 µm is Ideal for Use in Air/Gas and Vent Applications with Proprietary, Low Pressure Drop Hydrophobic PVDF Membrane., Pall Acro 50 In Line Vent Filter with Emflon II Membrane, Hose barb, no vent, 0.2 µm (100/pkg)
“Pall Laboratory products are designed for laboratory applications only. This product is not approved for use in medical, clinical, surgical or other patient applications. If you require further assistance in product selection for your chosen application, please contact Labcustomersupport@pall.com or call 1-800-521-1520."
“Pall Laboratory products are designed for laboratory applications only. This product is not approved for use in medical, clinical, surgical or other patient applications. If you require further assistance in product selection for your chosen application, please contact Labcustomersupport@pall.com or call 1-800-521-1520."
Specifications
Materials of Construction
- Filter Media: Emflon II membrane (hydrophobic PVDF)
- Housing: Polypropylene
Effective Filtration Area
- 20 cm2
Inlet/Outlet Connections
- Stepped hose barbs, 6.4 - 12.7 mm (1/4 - 1/2 in.) diameter with slip luer ID in the hose barb
Maximum Operating Temperature
- 60 °C (140 °F)
Maximum Operating Pressure
- 4.1 bar (410 kPa, 60 psi) at ambient temperature
Recommended Integrity Test Minimum Bubble Point - 60% IPA/40% Water (v:v)
- 1.1 bar (110 kPa, 16 psi)
Typical Air Flow Rate
- 27 L/min at 1.0 bar (100 kPa, 15 psi)
Endotoxin Level
- < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety
- Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Bacterial Retention
- Lot samples retain a minimum of 107 cfu/cm2 of B. diminuta per modified ASTM F838, current revision
Sterilization
- Provided non-sterile. Stable with gamma irradiation up to 50 kilogray. Can be sterilized by autoclaving at 121 °C (250 °F) for 20 min. Maximum autoclave stability, three cycles at 131 °C (268 °F) for 30 min.
Applications
- Venting of bioreactors, fermentation tanks, and carboys
- Gas purging of culture vessels
- Ideal for attachment to disposable systems that need to undergo gamma irradiation
Type
Vent Filters
Use
Sterile Filtration and Clarification
Related Products
Integrity Test Kit
- Verifies device integrity using one minute water breakthrough or bubble point tests
- Non-destructive test performed either in situ or in the lab
- For use with Acro 50 vent devices