Fluorodyne® II DJL Cartridges product photo Primary L

Fluorodyne® II DJL Cartridges

These 0.1 micron-rated filters with serial layer (0.2/0.1 micron) membrane construction assure high flow rates compared with other 0.1 micron (and even some 0.2 micron)-rated filters. Fluorodyne II grade DJL filters are validated for removal of Acholeplasma laidlawii mycoplasma (ATCC 23206) with a typical TR >108 and retention of Brevundimonas diminuta (ATCC 19146) at 107 CFU/cm² effective filter area (EFA). This allows for enhanced sterilization assurance as well as efficient mycoplasma control.

Pall’s standard AB-style filters are replaceable cartridges used with stainless steel housing assemblies, suitable for medium-to-large-scale manufacturing operations (100 L to > 1000 L). With Code 7 fittings as standard, these cartridges are the most versatile filtration units in Pall’s product range, and are available with a broader range of filtration media than any other filter type. These cartridges are also available encapsulated in the Kleenpak™ Nova capsule format.

Key Features and Benefits

  • Low extractables  
  • High protein transmission
  • Rapid preservative recoveries 
  • Easy integrity testing 
  • Compatible with organic solvents, acids and chemicals¹
  • Resin and surfactant-free 
  • Melt-sealed, non shedding
    ¹ Except ketones and amides

Quality Standards

  • 100% integrity tested
  • Manufactured for use in conformance with cGMP 
  • ISO 9000 Certified Quality System
  • Meets USP Biological Reactivity Test in vivo for Class VI-121 °C plastics
  • Every filter tested during manufacture
  • Test correlated to microbial retention
  • Certificate of Test provided includes
    • Fabrication integrity 
    • Bacterial retention 
    • Materials of Construction
    • Effluent quality for cleanliness, TOC and water conductivity, pH and pyrogens


* Required
Need Help?

Fluorodyne® II DJL membrane filter cartridge, 0.1 µm removal rating, 10-inch length, double-o-ring (silicone) with bayonet lock and fin end

Product ID: AB1DJL7PH4
Unit of Measure
1/EA
Min Order Qty
1
Removal Rating
0.1 µm
Membrane Material
Polyvinylidene fluoride (PVDF)
Membrane Brand
Fluorodyne II DJL
Industry-Specific Certification
Pharmaceutical Grade
Effective Filtration Area (Metric)
5500 cm²
Sterilization Option(s)
Suitable for Autoclave or SIP
See All Attributes

Fluorodyne® II DJL membrane filter cartridge, 0.1 µm removal rating, 5-inch length, double-o-ring (silicone) with bayonet lock and flat end

Product ID: AB05DJL2PH4
Unit of Measure
1/EA
Min Order Qty
1
Removal Rating
0.1 µm
Membrane Material
Polyvinylidene fluoride (PVDF)
Membrane Brand
Fluorodyne II DJL
Industry-Specific Certification
Pharmaceutical Grade
Effective Filtration Area (Metric)
2700 cm²
Sterilization Option(s)
Suitable for Autoclave or SIP
See All Attributes

Fluorodyne® II DJL membrane filter cartridge, 0.1 µm removal rating, 20-inch length, double-o-ring (silicone) with bayonet lock and fin end

Product ID: AB2DJL7PH4
Unit of Measure
1/EA
Min Order Qty
1
Removal Rating
0.1 µm
Membrane Material
Polyvinylidene fluoride (PVDF)
Membrane Brand
Fluorodyne II DJL
Industry-Specific Certification
Pharmaceutical Grade
Effective Filtration Area (Metric)
11000 cm²
Sterilization Option(s)
Suitable for Autoclave or SIP
See All Attributes
Description
These 0.1 micron-rated filters with serial layer (0.2/0.1 micron) membrane construction assure high flow rates compared with other 0.1 micron (and even some 0.2 micron)-rated filters. Fluorodyne II grade DJL filters are validated for removal of Acholeplasma laidlawii mycoplasma (ATCC 23206) with a typical TR >108 and retention of Brevundimonas diminuta (ATCC 19146) at 107 CFU/cm² effective filter area (EFA). This allows for enhanced sterilization assurance as well as efficient mycoplasma control.

Pall’s standard AB-style filters are replaceable cartridges used with stainless steel housing assemblies, suitable for medium-to-large-scale manufacturing operations (100 L to > 1000 L). With Code 7 fittings as standard, these cartridges are the most versatile filtration units in Pall’s product range, and are available with a broader range of filtration media than any other filter type. These cartridges are also available encapsulated in the Kleenpak™ Nova capsule format.

Key Features and Benefits

  • Low extractables  
  • High protein transmission
  • Rapid preservative recoveries 
  • Easy integrity testing 
  • Compatible with organic solvents, acids and chemicals¹
  • Resin and surfactant-free 
  • Melt-sealed, non shedding
    ¹ Except ketones and amides

Quality Standards

  • 100% integrity tested
  • Manufactured for use in conformance with cGMP 
  • ISO 9000 Certified Quality System
  • Meets USP Biological Reactivity Test in vivo for Class VI-121 °C plastics
  • Every filter tested during manufacture
  • Test correlated to microbial retention
  • Certificate of Test provided includes
    • Fabrication integrity 
    • Bacterial retention 
    • Materials of Construction
    • Effluent quality for cleanliness, TOC and water conductivity, pH and pyrogens
Specifications

Materials of Construction

 
 Filter Membrane  Hydrophilic modified PVDF
 Support/Drainage  Polypropylene
 Core/End Caps  Polypropylene
 Cage  Polypropylene with TiO2 whitener¹
 O-rings  Silicone elastomer
 Sealing Technology  Thermal bonding without adhesives
 
¹ TiO2 is an insoluble inorganic mineral filler that does not contribute to organic extractables

Operating Parameters²

 
 Maximum Temperature 80 ºC
 Maximum Differential Pressure (Forward Direction) 5.3 bar (77 psi) at 50 °C
3.4 bar (50 psi) at 80 °C
 Maximum Differential Pressure (Reverse Direction) 0.3 bar (4 psi)
 
² In compatible fluids which do not soften, swell, or adversely affect the filter or its materials of construction

Sterilization

 
 Autoclave and In Situ Steam 30 x 60 minutes at 125°C
10 x 60 minutes at 140°C
 

Typical Extractables in Water at 20 °C³

 
 < 5 mg per 254 mm (10 in.) cartridge
 
³  Tested on elements without pre-flushing

Nominal Effective Filter Area (EFA)

 
 5500 cm² (5.9 ft²) per 254 mm (10 in.) cartridge
 

Integrity Test Values (Air test gas, water wet)*

 
Max. allowable forward flow values
for 254 mm (10 in.) cartridge at 20°C
 29  mL/min at 4475 mbar (65 psi)
 
* Contact Pall for multi-element integrity test values and recommended test procedures

Typical Liquid Flow vs. Differential Pressure

Ordering Information
 
Item Part Number
Fluorodyne® II DJL Membrane Filter Cartridge
0.1 micron sterilizing grade and mycoplasma control, nominal length 125 mm (5 in.); Pall Code 7 double O-ring (silicone elastomer), bayonet lock and fin
 AB05DJL7PH4
Fluorodyne® II DJL Membrane Filter Cartridge
0.1 micron sterilizing grade and mycoplasma control, nominal length 125 mm (5 in.); Pall Code 2 double O-ring (silicone elastomer), bayonet lock, no fin
AB05DJL2PH4
Fluorodyne® II DJL Membrane Filter Cartridge
0.1 micron sterilizing grade and mycoplasma control, nominal length 254 mm (10 in.); Pall Code 7 double O-ring (silicone elastomer), bayonet lock and fin
AB1DJL7PH4
Fluorodyne® II DJL Membrane Filter Cartridge
0.1 micron sterilizing grade and mycoplasma control, nominal length 508 mm (20 in.); Pall Code 7 double O-ring (silicone elastomer), bayonet lock and fin
AB2DJL7PH4
Fluorodyne® II DJL Membrane Filter Cartridge
0.1 micron sterilizing grade and mycoplasma control, nominal length 762 mm (30 in.); Pall Code 7 double O-ring (silicone elastomer), bayonet lock and fin
AB3DJL7PH4
 

 

Application
Filters - Liquid - Mycoplasma Removal (0.1 micron)