PCCXS Series Component Cleanliness Cabinet product photo Primary L

PCCXS Series Component Cleanliness Cabinet

Please Note: The following information is provided to support existing users of Pall Cleanliness Cabinets (PCCs) for technical reference and order of consumables and spare parts. The complete cabinet assemblies are no longer available to purchase.


Description

The PCCXS Cleanliness Cabinet is the latest self-contained cleanliness cabinet from Pall, delivering the best practices in extracting particulate contamination from a component and retaining it on a test membrane for analysis.

The PCCXS series has been designed to assess cleanliness levels of small parts and components. It is the smallest unit of the PCC platform introduced by Pall.
Without standard, repeatable cleanliness validation, manufacturers and suppliers cannot meet industrial ISO standards.

Benefits include:

  • Provides an automated, repeatable process for checking parts cleanliness
  • Rapid to blank value* to start test sampling quickly
  • Risk of human error minimized
  • A fully HEPA filtered laminar air flow eliminates environmental cross-contamination
  • Test sample created is true representation of part contamination
  • Standard lab-friendly sized units to assess small components in accordance with ISO 18413, ISO16232 and VDA 19 procedures.
* relative value of cleanliness achieved over time, as specified by the customer

Features

  • Laminar air flow with 0.3 μm HEPA filter providing a controlled cleanliness environment (Class 5 per ISO 14644-1)
  • Easy to use, color touch screen human-machine interface using a tablet
  • Full work area access for service operation
  • Pressurized solvent dispensing and recycling circuits
  • Solvent vapor extracted by exhaust fan
  • Requires only a power source and exhaust vent




Contact Us
* Required

Documents

Datasheets

Download: English

Description

Please Note: The following information is provided to support existing users of Pall Cleanliness Cabinets (PCCs) for technical reference and order of consumables and spare parts. The complete cabinet assemblies are no longer available to purchase.


Description

The PCCXS Cleanliness Cabinet is the latest self-contained cleanliness cabinet from Pall, delivering the best practices in extracting particulate contamination from a component and retaining it on a test membrane for analysis.

The PCCXS series has been designed to assess cleanliness levels of small parts and components. It is the smallest unit of the PCC platform introduced by Pall.
Without standard, repeatable cleanliness validation, manufacturers and suppliers cannot meet industrial ISO standards.

Benefits include:

  • Provides an automated, repeatable process for checking parts cleanliness
  • Rapid to blank value* to start test sampling quickly
  • Risk of human error minimized
  • A fully HEPA filtered laminar air flow eliminates environmental cross-contamination
  • Test sample created is true representation of part contamination
  • Standard lab-friendly sized units to assess small components in accordance with ISO 18413, ISO16232 and VDA 19 procedures.
* relative value of cleanliness achieved over time, as specified by the customer

Features

  • Laminar air flow with 0.3 μm HEPA filter providing a controlled cleanliness environment (Class 5 per ISO 14644-1)
  • Easy to use, color touch screen human-machine interface using a tablet
  • Full work area access for service operation
  • Pressurized solvent dispensing and recycling circuits
  • Solvent vapor extracted by exhaust fan
  • Requires only a power source and exhaust vent


Specifications

Technical information

 
Overall Dimensions:
(W x D x H)
864 x 709 x 1927 mm
Working area:
(W x D x H)
600 x 600 x 500 mm
Sink 300 mm diameter
Weight: 175 kg (385 lb) approx
Materials: Working area:
Sink: Polished stainless steel 304L
Enclosure: Acrylic
Power supply: 230 V – 50 Hz, single phase
PLC: Proface
Power consumption 1.3 kW (without ultrasonics)
Reservoir (solvent): 13 L max (3.4 USg)
Rinsing flow rate:
(Adjustable)
5 L/min max.
(1.3 USgpm)
Solvent pen adjustable
Rinsing pressure: 2.8 bar max (41 psi)
 

The PCCXS series cabinets comply with the European Machinery Directive 2006/42/EC, Low voltage 2006/95/CE, and Electromagnetic compatibility 2004/108/CE and is fully CE compliant.

Technical information

Type

Diagnostic and Monitoring Equipment

Ordering Information

Pall Cleanliness Cabinet: PCCXS [1] [2] [3]

Table 1: Voltage Option

 
Code Description
1 110V / 50 & 60Hz, single phase
2 230V / 50 & 60 Hz, single phase
 
 

Table 2: Membrane Option

 
Code Description
S Single stage membrane holder
M 3 membrane multi-stage holder
 
 

Table 3: Ultrasonic Option

 
Code Description
OMIT No Ultrasonic transducer
US 100 100W Ultrasonic agitation power
 

Accessories

 
Code Description
PCCXSV2A62 Bowl for ultrasonic analysis (perforated bowl including mounting plate to be used during ultrasonic operation to hold test components)
PCCXSV2D60 Analysis mounting plate (perforated PEHD plate to hold test component during nozzle spraying)
PCCXSV2D4 Splashback guard (guard to avoid splashes during nozzle washing)
GHA0787OEM Cascade of 3 membranes
 

Analysis Membranes for Component Cleanliness Assessments

  • Ratings from 5 μm to 100 μm
  • Materials: Polyamide

see product datasheet M&EPCCMEMENa