PCM500 Series Fluid Cleanliness Monitor for Low Viscosity Applications Produktbild Primary L

PCM500 Series Fluid Cleanliness Monitor for Low Viscosity Applications

The PCM500 Fluid Cleanliness Monitor is a portable diagnostic monitoring device that provides a measurement of system fluid cleanliness.

Like the Pall PCM400, the new improved PCM500 uses proven mesh Blockage technology to report accurate, reliable, 3 part ISO 4406 cleanliness codes for most types of fluids, in many types of environment.

With the PCM500 you can:

  • Monitor contamination levels in mineral, synthetic, or water based fluids. Results are unaffected by the presence of water, air, or dark fluids
  • Get accurate, 3 part ISO 4406 cleanliness code results in under 6 minutes and quickly take preventative action
  • Upload real-time results directly to mobile devices for analysis and action
  • 'Pass off' cleanliness of new builds quickly and confidently
  • Protect your systems from catastrophic failure by detecting abnormal fluid cleanliness conditions quickly

The PCM500 can be permanently installed to monitor critical applications (including component test facilities) or used as a portable device for routine condition monitoring of various fluid systems

PCM500 Monitor Features

  • Proven mesh blockage technology provides accurate 3-part*  ISO 4406, AS 4059 Table 1 (NAS 1638) or AS 4059 Table 2 cleanliness codes
  • Self cleaning procedure between each test ensures optimum accuracy of results
  • Compact, robust, fully self contained portable design (fluid sampling pump included)
  • Simple to use, color touch screen interface
  • Long battery life for extended operation in remote locations
  • Measurement of fluid cleanliness, temperature, viscosity, and optional water content.
* 3 part code measured at 4 µm, 6 µm and 14 µm (c) per ISO 16889.

For ordering and pricing information contact your local sales office or distributor Contact Us.



* Required

Beschreibung

The PCM500 Fluid Cleanliness Monitor is a portable diagnostic monitoring device that provides a measurement of system fluid cleanliness.

Like the Pall PCM400, the new improved PCM500 uses proven mesh Blockage technology to report accurate, reliable, 3 part ISO 4406 cleanliness codes for most types of fluids, in many types of environment.

With the PCM500 you can:

  • Monitor contamination levels in mineral, synthetic, or water based fluids. Results are unaffected by the presence of water, air, or dark fluids
  • Get accurate, 3 part ISO 4406 cleanliness code results in under 6 minutes and quickly take preventative action
  • Upload real-time results directly to mobile devices for analysis and action
  • 'Pass off' cleanliness of new builds quickly and confidently
  • Protect your systems from catastrophic failure by detecting abnormal fluid cleanliness conditions quickly

The PCM500 can be permanently installed to monitor critical applications (including component test facilities) or used as a portable device for routine condition monitoring of various fluid systems

PCM500 Monitor Features

  • Proven mesh blockage technology provides accurate 3-part*  ISO 4406, AS 4059 Table 1 (NAS 1638) or AS 4059 Table 2 cleanliness codes
  • Self cleaning procedure between each test ensures optimum accuracy of results
  • Compact, robust, fully self contained portable design (fluid sampling pump included)
  • Simple to use, color touch screen interface
  • Long battery life for extended operation in remote locations
  • Measurement of fluid cleanliness, temperature, viscosity, and optional water content.
* 3 part code measured at 4 µm, 6 µm and 14 µm (c) per ISO 16889.

For ordering and pricing information contact your local sales office or distributor Contact Us.

Spezifikationenn

 
Power supply: 90-260 VAC or integral
12 VDC Lithium Ion battery
Battery life: Typically 35 samples
Temperature
Range:
10°C to 80°C (50°F to 176°F)
(dependent on fluid type)
Compatibility: Water glycols, aqueous solutions.
Petroleum and synthetic oils
(hydraulic lubricating, dielectric, etc.)
fuels, industrial phosphate esters, DI water
Seals: Nitrile
Operating viscosity: 0.5 to 40 cSt (29.5 to 185.3 SUS)
Pressure: 0 to 200 bar (2902 psi) max
Monitoring Range: ISO 4406: <11/9/7 to 23/21/17
SAE AS 4059 Table 1
Class 1 to 12 (derived from NAS 1638)
SAE AS 4059 Table 2
>4 µm 1A to 12A, >6 µm 1B to
12B >14 µm 1C to 12C
Accuracy: ± 1/2 ISO 4406 Code
Communication Ports: 3 x USB's (Data Acquisition, PC
Setup, Printer), Ethernet and RS232C (PLC Control)
Enclosure: IP 65 (NEMA 4)
Weight: 11 kg (24 lb)
Dimensions: 400 x 260 x 250 mm
(15.8 x 10.2 x 10 inches)
 


Real time data is displayed during test to show progress

Multiple test data can be stored and displayed for subsequent analysis and download

Typ

Diagnostic and Monitoring Equipment

Verwendung

Monitoring

Bestellinformation

Please select from the following part number options only.

 
PCM500 M A X357
PCM500 M B X357
PCM500 M D X357
PCM500 U B X357
PCM500 U C X357
 

Table 1: Fitting Type

 
Code Description
M 14" in. BSPP Female Swivel fitting to metric test point
U 14" in. NPT fitting and end cap
 

Table 2: Power Lead

 
Code Description
A UK Power Lead
B European Power Lead
C USA Power Lead
D Australian Power Lead
 

Printer Kit and Accessories

PCM500-PRT

For ordering and pricing information contact your local sales office or distributor Contact Us.

Rezensionen

Earn 10% off* your next order online by leaving a review of this product. Please login to your account to leave a review. We appreciate and value your feedback.

*Subject to Terms and Conditions.